IEC 60069-1 IEC Test Equipment Conductor Damage Degree Test Apparatus Rate 10 ± 2 R/Min
Contact Person : Tessa Huang
Phone Number : +86 18027403449
WhatsAPP : +8618027403449
Minimum Order Quantity : | 1 Set | Price : | Negotiatable |
---|---|---|---|
Packaging Details : | Plywood | Delivery Time : | 30 days |
Payment Terms : | T/T, L/C | Supply Ability : | 10 set per month |
Place of Origin: | China | Brand Name: | HeJin |
---|---|---|---|
Certification: | Third party calibration certificate/ CE(cost additional) | Model Number: | HJ0606 |
Detail Information |
|||
Product Name: | Oxygen Bomb (air Bomb) Ageing Tester HJ0606 | Usage: | Wire And Cable, Rubber |
---|---|---|---|
Material: | Stainless Steel | Temperature Range: | 0-200℃ |
Tank Volume: | 4000cm3 | Temperature Deviation: | ±1℃ |
Air Test Pressure: | 0.55 MPa ± 0.02 MPa | Test Gas: | Oxygen / Air |
Test Time: | 96h | Oxygen Test Pressure: | 2.1 MPa ± 0.07 MPa |
Highlight: | 0.55MPa IEC Test Equipment,4000cm3 IEC Test Equipment |
Product Description
IEC Test Equipment Ageing In An Air Bomb/Oxygen Bomb 0.55MPa Test Pressure
Standard:
IEC 60811-11-2 ‘Common test methods for insulating and sheathing materials of electric and optical cables — Part 1-2: General application — Thermal ageing methods’ clause 8.2 and 8.3 ;
IEC 60335-1 ‘Household and similar electrical appliances – Safety – Part 1: General requirements’ clause 22.32;
Application:
It is suitable for the aging test in the oxygen and compressed air to the wire and cable, rubber (plastic, leather) insulation and sheathing, as well as other polymer materials under the action of a predetermined pressure and temperature. IEC Test Equipment
Standard description:
The bomb shall be filled with air, which shall be free from oil and moisture, to a pressure of 0.55 ± 0.02 MPa. The test pieces shall be kept in the bomb at the temperature and for the time specified for the material in the standard for the type of cable.
The bomb shall be filled with commercial oxygen of not less than 97 % purity to a pressure of 2.1 ± 0.07 MPa. The test pieces shall be kept in the bomb at the temperature and for the time specified for the material in the standard for the type of cable.
Parameter:
Parameters Name | Parameter Data |
Power | AC 220V±10%, 50Hz |
Tank volume | 4000cm3 |
Heater power | 2.4KW |
Maximum Working Pressure | 4MP/cm2 |
Security Burst Pressure | 2.7-3.3MP/cm2 |
Temperature Deviation | ±1℃. IEC Test Equipment |
Temperature self-control range | Room temperature 0-200℃ |
Temperature control | Digital display automatic control |
Tank materials | Stainless steel |